All citations for Clove Lakes Health Care And Rehabilitation Center,

State inspectors check nursing homes that take Medicare or Medicaid against federal rules, usually about once a year, and also look into complaints. Each rule a home did not meet is a citation.

Back to Clove Lakes Health Care And Rehabilitation Center,

Last standard health inspection on record

October 16, 2024

This is more than 2 years ago. Newer inspection results may not be on record yet, so the results below may not describe the home today. Ask the home for its most recent inspection report.

Showing 1 to 11 of 11 citations

  1. No harm, with potential for more than minimal harm

    Have proper medical gas storage and administration areas.

    Pattern (more than a few residents). CMS level E.

    Fire safety citation K-0923, Gas, Vacuum, and Electrical Systems Deficiencies. Inspection of November 10, 2022. Found during a standard inspection.

    The home gave a date the problem would be corrected (January 9, 2023).

  2. No harm, with potential for more than minimal harm

    Meet requirements for outpatient facilities located next to inpatient facilities separated by fire resistive construction.

    Isolated (one or a few residents). CMS level D.

    Fire safety citation K-0132, Construction Deficiencies. Inspection of November 10, 2022. Found during a standard inspection.

    The home gave a date the problem would be corrected (January 9, 2023).

  3. No harm, with potential for more than minimal harm

    Install emergency lighting that can last at least 1 1/2 hours.

    Isolated (one or a few residents). CMS level D.

    Fire safety citation K-0291, Egress Deficiencies. Inspection of November 10, 2022. Found during a standard inspection.

    The home gave a date the problem would be corrected (January 9, 2023).

  4. No harm, with potential for more than minimal harm

    Have an enclosure around a vertical opening shaft.

    Isolated (one or a few residents). CMS level D.

    Fire safety citation K-0311, Smoke Deficiencies. Inspection of November 10, 2022. Found during a standard inspection.

    The home gave a date the problem would be corrected (January 9, 2023).

  5. No harm, with potential for more than minimal harm

    Ensure that special areas are constructed so that walls can resist fire for one hour or have an approved fire extinguishing system.

    Isolated (one or a few residents). CMS level D.

    Fire safety citation K-0321, Smoke Deficiencies. Inspection of November 10, 2022. Found during a standard inspection.

    The home gave a date the problem would be corrected (January 9, 2023).

  6. No harm, with potential for more than minimal harm

    Have properly sized and located compartments to protect residents from smoke.

    Isolated (one or a few residents). CMS level D.

    Fire safety citation K-0371, Smoke Deficiencies. Inspection of November 10, 2022. Found during a standard inspection.

    The home gave a date the problem would be corrected (January 9, 2023).

  7. No harm, with potential for more than minimal harm

    Have posted "No-smoking" signs in areas where smoking is not permitted or ashtrays provided where smoking was allowed.

    Isolated (one or a few residents). CMS level D.

    Fire safety citation K-0741, Miscellaneous Deficiencies. Inspection of November 10, 2022. Found during a standard inspection.

    The home gave a date the problem would be corrected (January 9, 2023).

  8. No harm, with potential for more than minimal harm

    Have a battery powered remote alarm panel in a location accessible by operating personnel.

    Isolated (one or a few residents). CMS level D.

    Fire safety citation K-0916, Gas, Vacuum, and Electrical Systems Deficiencies. Inspection of November 10, 2022. Found during a standard inspection.

    The home gave a date the problem would be corrected (January 9, 2023).

  9. No harm, with potential for minimal harm

    Keep aisles, corridors, and exits free of obstruction in case of emergency.

    Widespread (many residents). CMS level C.

    Fire safety citation K-0211, Egress Deficiencies. Inspection of November 10, 2022. Found during a standard inspection.

    The home gave a date the problem would be corrected (January 9, 2023).

  10. No harm, with potential for minimal harm

    Inspect, test, and maintain automatic sprinkler systems.

    Widespread (many residents). CMS level C.

    Fire safety citation K-0353, Smoke Deficiencies. Inspection of November 10, 2022. Found during a standard inspection.

    The home gave a date the problem would be corrected (January 9, 2023).

  11. No harm, with potential for minimal harm

    Have generator or other power source capable of supplying service within 10 seconds.

    Widespread (many residents). CMS level C.

    Fire safety citation K-0918, Gas, Vacuum, and Electrical Systems Deficiencies. Inspection of November 10, 2022. Found during a standard inspection.

    The home gave a date the problem would be corrected (January 9, 2023).

Source: CMS Provider Information, Fire Safety Deficiencies, Health Deficiencies, Citation Code Look-up, data as of September 30, 2026. Inspection results describe the home at the time of each inspection.