All citations for Wesbury United Methodist Commu

State inspectors check nursing homes that take Medicare or Medicaid against federal rules, usually about once a year, and also look into complaints. Each rule a home did not meet is a citation.

Back to Wesbury United Methodist Commu

Last standard health inspection on record

July 23, 2026

Showing 26 to 33 of 33 citations

  1. No harm, with potential for minimal harm

    Satisfy building requirements after a repair, renovation, modification, or change of user/occupancy.

    Pattern (more than a few residents). CMS level B.

    Fire safety citation K-0111, Construction Deficiencies. Inspection of August 16, 2024. Found during a standard inspection.

    The home gave a date the problem would be corrected (October 17, 2024).

  2. No harm, with potential for minimal harm

    Add doors in an exit area that do not require the use of a key from the exit side unless in case of special locking arrangements.

    Pattern (more than a few residents). CMS level B.

    Fire safety citation K-0222, Egress Deficiencies. Inspection of August 16, 2024. Found during a standard inspection.

    The home gave a date the problem would be corrected (October 17, 2024).

  3. No harm, with potential for minimal harm

    Provide exit doors that are held open by devices that will automatically close on the activation of a fire alarm or smoke detector.

    Pattern (more than a few residents). CMS level B.

    Fire safety citation K-0223, Egress Deficiencies. Inspection of August 16, 2024. Found during a standard inspection.

    The home gave a date the problem would be corrected (October 17, 2024).

  4. No harm, with potential for minimal harm

    Have properly located and lighted "Exit" signs.

    Pattern (more than a few residents). CMS level B.

    Fire safety citation K-0293, Egress Deficiencies. Inspection of August 16, 2024. Found during a standard inspection.

    The home gave a date the problem would be corrected (October 17, 2024).

  5. No harm, with potential for minimal harm

    Have properly installed hallway dispensers for alcohol-based hand rub.

    Pattern (more than a few residents). CMS level B.

    Fire safety citation K-0325, Smoke Deficiencies. Inspection of August 16, 2024. Found during a standard inspection.

    The home gave a date the problem would be corrected (October 17, 2024).

  6. No harm, with potential for minimal harm

    Ensure proper usage of power strips and extension cords.

    Pattern (more than a few residents). CMS level B.

    Fire safety citation K-0920, Gas, Vacuum, and Electrical Systems Deficiencies. Inspection of August 16, 2024. Found during a standard inspection.

    The home gave a date the problem would be corrected (October 17, 2024).

  7. No harm, with potential for minimal harm

    Ensure smoke barriers are constructed to a 1 hour fire resistance rating.

    Pattern (more than a few residents). CMS level B.

    Fire safety citation K-0372, Smoke Deficiencies. Inspection of October 27, 2023. Found during a standard inspection.

    The home gave a date the problem would be corrected (November 24, 2023).

  8. No harm, with potential for minimal harm

    Have simulated fire drills held at unexpected times.

    Pattern (more than a few residents). CMS level B.

    Fire safety citation K-0712, Miscellaneous Deficiencies. Inspection of October 27, 2023. Found during a standard inspection.

    The home gave a date the problem would be corrected (November 24, 2023).

Source: CMS Provider Information, Fire Safety Deficiencies, Health Deficiencies, Citation Code Look-up, data as of October 1, 2026. Inspection results describe the home at the time of each inspection.