All citations for Aurelia Osborn Fox Memorial Hospital

State inspectors check nursing homes that take Medicare or Medicaid against federal rules, usually about once a year, and also look into complaints. Each rule a home did not meet is a citation.

Back to Aurelia Osborn Fox Memorial Hospital

Last standard health inspection on record

April 22, 2025

Showing 1 to 21 of 21 citations

  1. No harm, with potential for more than minimal harm

    Conduct risk assessment and an All-Hazards approach.

    Widespread (many residents). CMS level F.

    Emergency preparedness citation E-0006, Emergency Preparedness Deficiencies. Inspection of April 22, 2025. Found during a standard inspection.

    The home gave a date the problem would be corrected (June 8, 2025).

  2. No harm, with potential for more than minimal harm

    Develop a communication plan.

    Widespread (many residents). CMS level F.

    Emergency preparedness citation E-0029, Emergency Preparedness Deficiencies. Inspection of April 22, 2025. Found during a standard inspection.

    The home gave a date the problem would be corrected (June 8, 2025).

  3. No harm, with potential for more than minimal harm

    Establish staff and initial training requirements.

    Widespread (many residents). CMS level F.

    Emergency preparedness citation E-0037, Emergency Preparedness Deficiencies. Inspection of April 22, 2025. Found during a standard inspection.

    The home gave a date the problem would be corrected (June 8, 2025).

  4. No harm, with potential for more than minimal harm

    Have exits that are accessible at all times.

    Widespread (many residents). CMS level F.

    Fire safety citation K-0271, Egress Deficiencies. Inspection of April 22, 2025. Found during a standard inspection.

    The home gave a date the problem would be corrected (May 16, 2025).

  5. No harm, with potential for more than minimal harm

    Install proper backup exit lighting.

    Widespread (many residents). CMS level F.

    Fire safety citation K-0281, Egress Deficiencies. Inspection of April 22, 2025. Found during a standard inspection.

    The home gave a date the problem would be corrected (June 14, 2025).

  6. No harm, with potential for more than minimal harm

    To conduct inspection, testing and maintenance of fire doors by qualified individuals.

    Widespread (many residents). CMS level F.

    Fire safety citation K-0761, Miscellaneous Deficiencies. Inspection of April 22, 2025. Found during a standard inspection.

    The home gave a date the problem would be corrected (May 22, 2025).

  7. No harm, with potential for more than minimal harm

    Ensure that building systems meet requirements determined by risk assessment procedures performed by qualified personnel.

    Widespread (many residents). CMS level F.

    Fire safety citation K-0901, Electrical Deficiencies. Inspection of April 22, 2025. Found during a standard inspection.

    The home gave a date the problem would be corrected (May 22, 2025).

  8. No harm, with potential for more than minimal harm

    Establish staff and initial training requirements.

    Widespread (many residents). CMS level F.

    Emergency preparedness citation E-0037, Emergency Preparedness Deficiencies. Inspection of May 26, 2022. Found during a standard inspection.

    The home gave a date the problem would be corrected (July 15, 2022).

  9. No harm, with potential for more than minimal harm

    Have generator or other power source capable of supplying service within 10 seconds.

    Widespread (many residents). CMS level F.

    Fire safety citation K-0918, Gas, Vacuum, and Electrical Systems Deficiencies. Inspection of May 26, 2022. Found during a standard inspection.

    The home gave a date the problem would be corrected (July 15, 2022).

  10. No harm, with potential for more than minimal harm

    Ensure receptacles at patient bed locations and where general anesthesia is administered, are tested after initial installation, replacement or servicing.

    Widespread (many residents). CMS level F.

    Fire safety citation K-0914, Gas, Vacuum, and Electrical Systems Deficiencies. Inspection of October 31, 2019. Found during a standard inspection.

    The home gave a date the problem would be corrected (December 30, 2019).

  11. No harm, with potential for more than minimal harm

    Ensure smoke barriers are constructed to a 1 hour fire resistance rating.

    Pattern (more than a few residents). CMS level E.

    Fire safety citation K-0372, Smoke Deficiencies. Inspection of April 22, 2025. Found during a standard inspection.

    The home gave a date the problem would be corrected (May 22, 2025).

  12. No harm, with potential for more than minimal harm

    Have properly installed electrical wiring and gas equipment.

    Pattern (more than a few residents). CMS level E.

    Fire safety citation K-0511, Services Deficiencies. Inspection of April 22, 2025. Found during a standard inspection.

    The home gave a date the problem would be corrected (May 22, 2025).

  13. No harm, with potential for more than minimal harm

    Have a battery powered remote alarm panel in a location accessible by operating personnel.

    Pattern (more than a few residents). CMS level E.

    Fire safety citation K-0916, Gas, Vacuum, and Electrical Systems Deficiencies. Inspection of May 26, 2022. Found during a standard inspection.

    The home gave a date the problem would be corrected (July 15, 2022).

  14. No harm, with potential for more than minimal harm

    Have an enclosure around a vertical opening shaft.

    Pattern (more than a few residents). CMS level E.

    Fire safety citation K-0311, Smoke Deficiencies. Inspection of October 31, 2019. Found during a standard inspection.

    The home gave a date the problem would be corrected (January 16, 2020).

  15. No harm, with potential for more than minimal harm

    Have approved installation, maintenance and testing program for fire alarm systems.

    Pattern (more than a few residents). CMS level E.

    Fire safety citation K-0345, Smoke Deficiencies. Inspection of October 31, 2019. Found during a standard inspection.

    The home gave a date the problem would be corrected (December 30, 2019).

  16. No harm, with potential for more than minimal harm

    Ensure smoke barriers are constructed to a 1 hour fire resistance rating.

    Pattern (more than a few residents). CMS level E.

    Fire safety citation K-0372, Smoke Deficiencies. Inspection of October 31, 2019. Found during a standard inspection.

    The home gave a date the problem would be corrected (December 30, 2019).

  17. No harm, with potential for more than minimal harm

    Have properly installed electrical wiring and gas equipment.

    Pattern (more than a few residents). CMS level E.

    Fire safety citation K-0511, Services Deficiencies. Inspection of October 31, 2019. Found during a standard inspection.

    The home gave a date the problem would be corrected (December 30, 2019).

  18. No harm, with potential for more than minimal harm

    To conduct inspection, testing and maintenance of fire doors by qualified individuals.

    Pattern (more than a few residents). CMS level E.

    Fire safety citation K-0761, Miscellaneous Deficiencies. Inspection of October 31, 2019. Found during a standard inspection.

    The home gave a date the problem would be corrected (January 7, 2020).

  19. No harm, with potential for more than minimal harm

    Ensure that testing and maintenance of electrical equipment is performed.

    Pattern (more than a few residents). CMS level E.

    Fire safety citation K-0921, Gas, Vacuum, and Electrical Systems Deficiencies. Inspection of October 31, 2019. Found during a standard inspection.

    The home gave a date the problem would be corrected (December 30, 2019).

  20. No harm, with potential for more than minimal harm

    Install corridor and hallway doors that block smoke.

    Isolated (one or a few residents). CMS level D.

    Fire safety citation K-0363, Smoke Deficiencies. Inspection of April 22, 2025. Found during a standard inspection.

    The home gave a date the problem would be corrected (May 21, 2025).

  21. No harm, with potential for more than minimal harm

    Ensure that personnel concerned with handling of medical gases and cylinders are trained on the risk.

    Isolated (one or a few residents). CMS level D.

    Fire safety citation K-0926, Gas, Vacuum, and Electrical Systems Deficiencies. Inspection of October 31, 2019. Found during a standard inspection.

    The home gave a date the problem would be corrected (December 30, 2019).

Source: CMS Provider Information, Fire Safety Deficiencies, Health Deficiencies, Citation Code Look-up, data as of October 1, 2026. Inspection results describe the home at the time of each inspection.